Scenario Playbooks
Use these playbooks as practical workflows. Your tenant labels may differ slightly, but the states and decisions are the same.
Scenario 1: Routine Client Sample#
- Open the client.
- Add a sample with the correct contact, sample type, date sampled, and profile.
- Save and print labels if needed.
- Receive the sample when it arrives.
- Assign analyses to a worksheet or enter directly.
- Submit results.
- Verify.
- Publish the COA.
Scenario 2: Physical Sample Arrives Before Registration#
- Inspect the sample and COC.
- Create the client or contact if missing.
- Register the sample using the physical collection details.
- Receive immediately, or rely on auto-receive if enabled.
- Attach the COC scan.
Scenario 3: Out-of-Range Routine Result#
- Save the result to trigger range checks.
- Review the specification flag.
- Check QC, instrument status, sample condition, and calculations.
- Add interpretation or remarks if required.
- Submit only when lab SOP allows.
- The verifier decides whether to verify, retract, rerun, or reject.
Scenario 4: Result Needs Retesting Before Verification#
- The verifier opens the submitted sample or worksheet.
- Select the affected analysis.
- Use Retract and add a reason.
- The analyst corrects or reruns the result.
- The analyst submits again.
- The verifier reviews and verifies.
Scenario 5: Published Report Had an Error#
- Do not edit silently.
- Notify the lab manager.
- Locate the published report and audit trail.
- Follow the invalidation or amendment process.
- Document the reason, corrected value, and client notification.
- Publish the amended report according to lab policy.
Scenario 6: Batch Bench Work#
- Receive all samples.
- Create a worksheet from a template.
- Add unassigned received analyses.
- Add QC controls, blanks, or duplicates.
- Enter results.
- Submit the worksheet.
- The verifier reviews routine and QC data together.
Scenario 7: New Test Requested#
- Do not create an ad hoc result field on a live sample unless authorized.
- Send support the test name, method, unit, detection limits, precision, category, price if used, and report expectations.
- Clearline configures or updates the analysis service.
- Add the service to profiles or templates if it should be routine.
Scenario 8: Client Needs Portal Access#
- Confirm the contact belongs to the correct client.
- Confirm the email address.
- Request portal access from support.
- After access is granted, the contact can view only their organization's scoped data.